<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>22112</ErrorNumber>
<ErrorFamily>Autoloader 2</ErrorFamily>
<ErrorMessage>
ALE2: Drop Plate onto table sensor, Push the START Button to finish the plate unload
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>ThermoFlex</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
This is only a diagnostic message.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
As described in the message.
The goal here is to let the plate lay on the table so that it covers the plate
length sensor and then press the START button to have the plate driven out of the
machine by the drum.
</text></DescriptionText>
</OperatorAction>
<Cause>
The machine has finished jogging the plate far enough for it to be laid on the table
and is now asking for the plate to be let down.
</Cause>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:33:56 2008 PST
</Updated>
<Links>
</Links>
</Error>